Calibration of Pipette, Single Mark

Calibration of Pipette, Single Mark

Calibration of Pipette, Calibration of  Single Mark Pipette,

Calibration of Pipette, Single Mark

1.0 Purpose: This SOP describes a procedure for the measurement of capacity of single mark pipette.

2.0 Scope: This SOP is applicable to all single mark pipettes of different nominal capacity.

3.0 Responsibility: It is the responsibility of Quality Assurance Manager to ensure that the procedure is followed.

4.0 Materials and Equipment Used:

4.01 A weighing balance that is calibrated with traceable weights and regularly maintained.

4.02 A calibrated thermometer.

4.03 A weighing vessel.

4.04 Distilled water.

4.05 Acetone.

4.06 4% Acetic acid solution

4.07 Worksheets or forms to record the raw data.

5.0 Procedure: The determination of capacity of pipettes is based on weighing water and calculating mass and volume based on the density at the appropriate temperature.

5.01 Clean the pipette with soap solution and water to remove any adhering dirt or grease. Rinse the pipette with acetic acid solution and then wash with water till free from all acid.

5.02 Dry the pipette by rinsing with acetone.

5.03 Weigh a cleaned and dried weighing vessel (W1).

5.04 Fill the pipette with distilled water at ambient temperature to a few ml above the graduation line.

5.05 Then hold the pipette in a vertical position, and adjust the falling meniscus to the graduation line as given below –

(a) Set the meniscus so that the plane of the top edge of the graduation line is horizontally tangential to the lowest point of the meniscus, the line of sight being in the same plane.

(b) Remove any drop adhering to the jet of the pipette by bringing the surface of a glass vessel into contact with the tip of the jet.

5.06 Deliver the water thus measured into cleaned tared vessel slightly inclined so that the tip of the jet   is in contact with the inside of the vessel, but without movement of the one against the other throughout the delivery and waiting periods.

5.07 Allow the pipette to empty till the meniscus comes to rest in the jet.

5.08 Again weigh the vessel containing distilled water (W2).

5.09 Note the temperature of water and calculate the volume of water delivered by the pipette at the reference temperature of 270C.

6.0 Calculation: Calculate the volume of water delivered by pipette using the following formula –

V= W2 – W1 / D

Where:

V    = Volume of the pipette, Cm3 (ml )

W1 = Weight of empty weighing vessel.

W2 = Weight of empty weighing vessel & distilled water.

D    = Density of water at 270C (i.e. 0.9965 g /cc)

7.0 Reporting: Record all the observations in the worksheets or forms. Calculate the correct volume of water delivered by the pipette and, the amount by which this differs from the indicated volume, is the volumetric error.

Also, Visit: