SOP for calibration of Auto Titrator

SOP for calibration of Auto Titrator

1.0
  PURPOSE 

        The purpose of the standard operating procedure is
to provide guidelines for calibration of Auto Titrator
.

2.0   SCOPE

        The
scope of this standard operating procedure includes calibration of
Auto Titrator.

3.0    RESPONSIBILITY

3.1    Section In charge

3.2    Technical Manager

3.3    Q.A. Manager

4.0    PROCEDURE

4.1    DETERMINATION OF VOLUME:

4.1.1 Clean the measuring / dispensing
pipette available with the respective set of the,    

           equipment.

4.1.2   Fill the measuring pipette with distilled
water and adjust to the zero point.

4.1.3  
Dispense the known volumes of water at random or from the set point of
the   

           graduation marks of the pipette.

4.1.4   
Repeat the dispensing of water at different set points and weigh the
beaker again

and again.

4.1.5
  From the different weights obtained
calculate the volume dispensed and co-relate

with
the graduation of the pipette and assign the value.

4.2     CALCULATION:

                                                             Weight of the water dispensed (gm)

     Volume of the pipette (ml),  V=       ————————————————                    

                                                             Density of Water at 27oC (gm / ml)

       Where,

               
D = Density (gm/ml) of Water at 27oC