SOP on Calibration of Photofluorimeter

 Calibration of Photofluorimeter

SOP on Calibration of Photofluorimeter

SOP on Calibration of Photofluorimeter

1.0 PURPOSE   

The purpose of the standard operating procedure is to provide guidelines for calibration of Photofluorimeter.

2.0 SCOPE  

The scope of this standard operating procedure includes the calibration of Photofluorimeter.

3.0 RESPONSIBILITY

3.1 Section In charge

3.2 Technical Manager

3.3 Q.A. Manager

4.0 PROCEDURE

4.1 On the instrument 15 minutes before starting the calibration.

4.2 Make dilutions using fluoroscene in 0.1 M NaOH (1ppm, 2ppm, 3ppm and 4 ppm).

4.3 Use 0.1M NaOH as blank and set the instrument on “0 “with blank knob.

4.4 Set the instrument at ‘20’ with 1ppm solution.

4.5 Take the readings at 1, 2, 3 and 4 ppm without touching any of the knobs.

4.6 Standard readings for above mentioned concentrations:

S. No.

Concentration(ppm)

Observed Intensity(Units)

1

1 ppm

21

2

2 ppm

40

3

3 ppm

60

4

4 ppm

79

 

 