SOP on Calibration of BOD Incubator

Calibration of BOD Incubator

SOP on Calibration of BOD Incubator

SOP on Calibration of BOD Incubator

1.0 PURPOSE

The purpose of the standard operating procedure is to provide guidelines for calibration of BOD Incubator.

2.0 SCOPE

The scope of this standard operating procedure includes calibration of BOD Incubator.

3.0 RESPONSIBILITY

3.1Section Incharge

3.2 Technical Manager

3.3 Q.A. Manager

4.0 PROCEDURE

4.1 Switch on the instrument.

4.2 Ensure that the temperature indicator/controller is in working condition i.e. the set temperature should be maintained/retained properly.

4.3 Ensure that the door of incubator is properly closed and does not loosen during operation.

4.4 Ascertain and set the points/points at which calibration is required.

4.5 Switch on the incubator and allow it to reach the set point temperature.

4.6 Apparatus Required For Calibration: Platinum resistance thermometer with sensor and digital temperature indicator.

4.7 CALIBRATION:

4.7.1 Insert the sensor of the reference/calibrator sensor (PRT) in the BOD incubator as near as possible to the sensing tip or sensor of the temperature controller/indicator of the BOD incubator.

4.7.2 When the temperature is stable note down the temperature reading as indicated by the BOD incubator and the reference indicator simultaneously.

4.7.3 Set the next point and record the temperature readings accordingly.

4.7.4 Take at least three to four readings for better accuracy and reproducibility.

4.7.5 Correlate the readings observed with the reference standards and apply corrections, if necessary.

Also, Visit:

B. Pharma Notes | B. Pharma Notes | Study material Bachelor of Pharmacy pdf

B. Pharma Handwritten Notes

B. Pharma PDF Books

B. Pharma Lab Manual

D. Pharma Lab Manual

B. Pharma 8th Semester Previous Year Question Paper

D. Pharma Notes